DRAFT
December 16, 2004 3:54 pm, GSG_Chapter_4-02.fm
Chapter 2 Designing Primers and Probes for Quantification Assays
Manually Designing Primers and Probes
14 Primer Express Software 3.0 Getting Started Guide
Notes
Figure 4. Primer Probe Test Tool dialog box
4. From the Document Type drop down menu, select the desired document type.
Verify that the Parameter box is set to Default. For more information about
changing parameters, see Primer Express Software Version 3.0 Online Help.
5. Paste (Ctrl+V) the putative sequence in the Probe 1 field. The Primer Probe Test
Tool displays the Tm, %GC, and the oligonucleotide length to the right of the Probe
1 field.
6. If the Tm is not between 68 °C to 70 °C, highlight a section of the sequence to view
the corresponding Tm, %GC, and oligonucleotide length. Once the highlighted
region results in the desired Tm, click on Trim to delete the non-highlighted bases.
Ensure the following guidelines are met (for more information on design
guidelines, refer to Primer Express Software Version 3.0 Online Help):
Amplicon Length – 50 to 150 bases for optimum PCR efficiency.
Probe Length – 13 to 25 bases (13 to 30 bases if using conventional
TaqMan probes)
Tm – 68 °C to 70 °C.
% GC – 30% to 80%.