20. CHANGING THE SETTINGS
110
& Prism constant correction
Reflective prisms each have their prism constant.
Set the prism constant correction value of the reflective prism you are using. When selecting "N-
Prism" in "Reflector", prism constant correction value is set to "0" automatically.
I "ppm" tab
[0ppm]: Atmospheric correction factor returns to 0 and temperature and air pressure are set to
the factory settings.
Atmospheric correction factor is calculated and set using the entered values of the temperature
and air pressure. Atmospheric correction factor can also be entered directly.
Wnen the settings here are different between basic mode and program mode, program mode
settings are prior for use.
Items set, options, and input range (*: Factory setting)
Temperature: -30 to 60°C (15*), -22 to +140°F (59*)
Pressure: 500 to 1400hPa (1013*), 375 to 1050mmHg (760*),
14.8 to 41.3 (inchHg) (29.9*)
Humidity: 0 to 100% (50*)
ppm (Atmospheric correction factor): -499 to 499 (0*)
•The "Humidity" item is displayed only when the "ppm setting" in "Obs. condition" is set to
"+Humidity".